Analyst® Software Scheduled MRM Acquisition Failure Due To Unprogrammed Diverter Valve


Date: 03/14/2018
Categories: Analyst Software

0 Votes
   Print    Rate Article:

If a scheduled MRM (sMRM) acquisition methods created in Analyst® software fails to start, then it is very likely that the diverter valve doesn't have a time table programmed in the acquisition method. If the diverter valve is not required in your method, then right-click on the valve icon, unselect the Use option and save the method. Alternatively, the time table can be programmed for the divert valve to allow the sMRM method to acquire.