Adding User Groups to the Analyst® Software Secruity Configuration


Date: 03/08/2018
Categories: Analyst Software

0 Votes
   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Answer

To enable users to launch Analyst® software via a Windows user group, the group must be added to the Security Configuration.

In Analyst software, open the security configuration dialog box. Ensure that the security mode is set to either Integrated or Mixed-mode. In the People tab, select the New Person radio button. Associate each windows user group with the corresponding Analyst software role by selecting the windows user and clicking the "Add button" to add the appropriate role.
User-added image