I am having trouble reaching temperature in the ALS Thermostat. What is the effective range of the ACMP Peltier?


Date: 12/04/2020
Categories:

0 Votes
   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Answer

The ACMP Peltier effectivity is based on the ambient temperature of the instrument.  The operational range of the ACMP is between 4-40C.  In order to reach 4C, the room temperature needs to be 30C or lower with a relative humidity of less than 70%.  However, the temperature accuracy is +/-6C.