Peak Labels from Analyst® Report Are Missing in MultiQuant™ Software Report


Date: 10/19/2018
Categories: MultiQuant Software

0 Votes
   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Issue Description

In Analyst® software, the peaks showing in the XIC automatically had peak labels showing in Analyst reports for each analyte. Those types of labels are not automatically added by MultiQuant™ software.


Resolution

To add in the labels to reports generated in MultiQuant software, the tag for Peak Review has to be edited to show the Analyte_Peak_Name in the Label field.