Make the Analytics workspace in SCIEX OS more user-friendly


Compared to Analyst Quantitate / MultiQuant, routine customers miss the following functionalities: - Copy/paste of graphs - display of navigation controls by default - possibility to maximize results table and calibration curve pane - after activating the manual integration, the user first needs to click into the pane before he can draw the manual baseline. In MQ, it is directls possible without the need to click into the graph first. Moreover, after drawing the baseline manually, the scroll function of the mouse is inactive, the user needs to click into the next graph before it is active again - when reviewing the chromatograms of a group, the user would like to automatically pass over to the next group. This is only possible using a work-around. In MQ the display changed directly and went over to the next group if the user uses the scroll funtion of the mouse or the arrows of the keyboard. - if the internal standard area is zero, the calculated concentration for the analyte is displayed as N/A. The user may eventually not realize that the sample is in fact positive. In MW, a # symbol was shown in this case, which makes it easier for the user to see that something is wrong and that he should review the results more carefully.


0 Comments Jan 17, 2020 119 Views Product: SCIEX OS Software