Answer
In a SCIEX OS qmethod component specific values (or text) can be stored.
In actual versions you can find a component column in the quantitation method - but you have to make it visible first:
Open the method, select <Components> in the menue on the left hand side. In <Options> open the <Table Settings> and check <Comment>:
An additional column will be visible in the components table and can be filled:
These values will be populated in the results table column [Component Comment].
This can be done vice versa, too: new/changed values in the [Component Column] in the results table will result in an updated quantitation method.
In earlier versions (prior to v2.2) you have to use this vice versa approach, as the <Comment> column in the component table of the quantitation method is not available yet.