Setting Up the LC Pressure Trace Monitoring for an Acquisition Method in Analyst® Software


Date: 03/03/2023
Categories: Software

0 Votes
   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Answer

Analyst® software 1.7.3 supports LC pump pressure trace and column oven temperature trace monitoring. The resulting pressure traces can be stored in the data file for each sample. 

To set up the LC pressure trace monitoring use the following steps:

1) In the Create New Hardware Profile or Edit Hardware Profile dialog, select the LC system and then click Setup Device.
2) Select the Enable Aux Trace Monitoring check box.

enable pressure trace.PNG
3) Make sure that the LC system configuration is complete, and then click OK.
4.) Activate the hardware profile.
5) Acquire data. The acquired data file (.wiff) will contain the LC traces.