Date: | 11/03/2021 |
Categories: | Analyst Software , SCIEX OS , MultiQuant Software |
For research use only. Not for use in diagnostic procedures.
Issue Description
When opening a template, an error was received from the Template Content Control Manager, Error: Specified argument was out of range of valid values.
Resolution
To resolve this error, the Template Content Control Manager needs to be reinstalled. To start the new installation, navigate to C:\Program Files\AB SCIEX\ReporterOfficeAddins\TemplateContentControlManager and then double click on the file TemplateContentControlManager.vsto. Once installed, the templates will be able to be edited.
Posted: now