Date: | 11/02/2021 |
Categories: | MultiQuant Software |
For research use only. Not for use in diagnostic procedures.
Issue Description
When using the reporter system inside of MultiQuant™ software, the chromatograms did not appear on the report after printing.
Resolution
If the Peak Fill option is set to "Do Not Fill Peaks" in MultiQuant software, chromatograms will be faded out when output to a report and sometimes appear as not legible. To make the chromatograms show up more pronounced, change the Fill Peak option to "Using Solid Style." To do this, right-click within Peak Review Pane, choose Option, under Appearance Tab find Fill Peak and select "Using Solid Style" from the drop-down menu. Click OK to save the change. When the report is created and printed, the filled chromatograms appear much more defined in the report.
Posted: now