Date: | 11/01/2021 |
Categories: | Analyst Software |
For research use only. Not for use in diagnostic procedures.
Issue Description
In Analyst® software 1.7.2, users cannot save or submit an acquisition batch if the project path is longer than 126 characters. When this type of file is submitted, the following error message: “Failed to set Header data” is shown. When this file is saved, the following error message: “Failed to save BatchName” is shown.
Resolution
There is no resolution for this issue yet. The fix will be considered in future releases.
Posted: now