Answer
PEEKSil tubing is normally used to connect a Waters ACQUITY M-Class UPLC to a ZenoTOF 7600 system. The PEEKSil tubing has an outer ID of 1/32 inch, but the tubing for the Waters M-Class has an outer ID of 1/16 inch, so purchasing the correct fittings are critical in order to connect LC components, such as trap columns, to the ZenoTOF 7600 system.
The recommended hardware for making the connection between the ZenoTOF 7600 system and the M-Class UPLC tubing are: