What Is the Minimum Accumulation Time in Analyst® TF Software for TripleTOF® Systems?


Date: 09/30/2022
Categories: Analyst Software

0 Votes
   Print    Rate Article:

For research use only. Not for use in diagnostic procedures.


Answer

The minimum accumulation time supported in Analyst® TF software is 50 ms for TOF MS experiments and 10 ms for TOF MS/MS experiments. For SWATH methods, it's recommended not to go lower than 100 ms for TOF MS and 20 ms for TOFMS/MS experiments.