There are four different multiple reaction monitoring (MRM) modes that can be acquired using SCIEX TripleQuad or QTRAP systems with SCIEX OS software.
1) MRM: All transitions are monitored for every cycle throughout the entire method duration.
2) Scheduled MRM: Transitions are only monitored within their defined retention time window during acquisition.
3) Scout triggered MRM RT mode: The marker transition and its dependent transitions are monitored for every cycle until the next marker transition exceeds its threshold.
4) Scout triggered MRM group mode: Transitions are monitored within their defined retention time window if the marker transition for the super group exceeds its threshold.
Scout triggered MRM is a new acquisition method type in SCIEX OS software version 3.1.6.